| Model Number |
IPN046779 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/12/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The device has not been returned for investigation.Teleflex will continue to monitor and trend related events.
|
| |
|
Event or Problem Description
|
|
Per the doctor catheter tip broke off when removing from patient and tip is still in patients epidural space.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).A lot number was not provided.A device history record review was performed based upon a lot number from sales history data.A device history record review was performed on the epidural catheter with no relevant findings.A corrective action is not required at this time as the root cause for this complaint investigation could not be determined based upon the information provided and without the actual sample.Complaint verification testing could not be performed as no sample was returned for analysis.A lot number was not provided.A device history record review was performed based upon a lot number from sales history data.A device history record review was performed on the epidural catheter with no relevant findings.Therefore, the potential cause of this complaint could not be determined based upon the information provided and without a sample.No further action is required at this time.
|
| |
|
Event or Problem Description
|
|
Per the doctor catheter tip broke off when removing from patient and tip is still in patients epidural space.
|
| |
|