| Model Number |
H700488 |
| Medical Device Problem Code |
Thermal Decomposition of Device (1071)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/06/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The results of the investigation are inconclusive as the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.No non-conformances associated with the reported event were identified.Based on the information received, the cause of the reported thermal event could not be determined.
|
| |
|
Event or Problem Description
|
|
Following the procedure, an outlet cable that was plugged into an outlet was connected to the generator grounding port and the metal inside became burnt.There were no adverse consequences to the user who connected the outlet cable into the generator grounding port.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d9, g3, h2, h3, h6.One ampere¿ rf ablation generator was received into the lab for analysis.Visual inspection of the returned ampere generator revealed a damage/marred indifferent electrode connector pin which would have led to the reported complaint.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.No non-conformances associated with the reported event were identified.Based on the information provided to abbott and the investigation performed, the reported event was confirmed and isolated to a damage/marred indifferent electrode connector pin.It is unknown how the damage occurred.
|
| |
|