• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RICHARD WOLF GMBH E-LINE; URETERORENOSCOPE 12° 8/9.8FR WL 430MM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RICHARD WOLF GMBH E-LINE; URETERORENOSCOPE 12° 8/9.8FR WL 430MM Back to Search Results
Model Number 8703.534
Medical Device Problem Codes Use of Device Problem (1670); Device Handling Problem (3265)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/02/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The instrument has not yet been sent for examination by the user.As soon as the examination is completed, we will send a follow-up report.
 
Event or Problem Description
It was reported to richard wolf "on (b)(6) 2021, fibrous ureteroscopic holmium laser lithotripsy was performed.In the process of lithotripsy, the fibrous ureteroscope failed to bend suddenly and kidney stones could not be found, so the operation was aborted.No information is known about the patient's state of health.".
 
Additional Manufacturer Narrative
Follow-up report #1 is to provide fda with missing information (results of the device evaluation/investigation), new information, and changed information.The following is the results of the device investigation/evaluation as reported by the manufacturer: using process: on (b)(6) 2021, fibrous ureteroscopic holmium laser lithotripsy was performed.In the process of lithotripsy, the fibrous ureteroscope failed to bend suddenly and kidney stones could not be found, so the operation was aborted.Detail of the investigation: after checking, this scope had problem during operation, so the operation was aborted.After that, hospital sent it for a third party company for repair, now it is in normal use.Analysis of cause: normal wear.Instructions for use: the hospital found a third party who carried out the repair himself.The instrument has signs of third-party repair, as a result of which it does not correspond to the specifications given by richard wolf gmbh and it can be ruled out that the instrument was delivered in this condition.This means that richard wolf is no longer the manufacturer of the product.All claims against richard wolf are extinguished by the third-party repair.Product risk: the hospital found a third party who carried out the repair himself.The instrument has signs of third-party repair, as a result of which it does not correspond to the specifications given by richard wolf gmbh and it can be ruled out that the instrument was delivered in this condition.This means that richard wolf is no longer the manufacturer of the product.All claims against richard wolf are extinguished by the third-party repair.Richard wolf gmbh considers this matter closed.However, in the event rwgmbh receives any additional information a follow up report will be submitted to fda.Richard wolf medical instruments corporation(rwmic) submitting report on behalf of rwgmbh report.
 
Event or Problem Description
Please see manufacturers narrative for device evaluation/investigation results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
E-LINE
Common Device Name
URETERORENOSCOPE 12° 8/9.8FR WL 430MM
Manufacturer (Section D)
RICHARD WOLF GMBH
pforzheimerstr. 32
knittlingen 75438
GM  75438
Manufacturer (Section G)
RICHARD WOLF GMBH
pforzheimerstr. 32
knittlingen 75438
GM   75438
Manufacturer Contact
heiko seider-biedermann
pforzheimerstr. 32
knittlingen 75438
GM   75438
MDR Report Key12705292
Report Number9611102-2021-00035
Device Sequence Number5511714
Product Code FGB
UDI-Device Identifier04055207018798
UDI-Public04055207018798
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K963855
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Physician
Remedial Action Replace
Type of Report Initial,Followup
Report Date (Section B) 01/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number8703.534
Device Catalogue Number8703.534
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/27/2021
Supplement Date Received by Manufacturer01/04/2022
Initial Report FDA Received Date10/27/2021
Supplement Report FDA Received Date01/05/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/19/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-