Model Number 1516-40-610
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Pain (1994); Joint Laxity (4526)
Date of Event 10/14/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
The patient had presented with pain and no implant appeared loose.The implant was somewhat unstable.Doi: 2015, dor: (b)(6) 2021, right knee.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination, therefore unavailable for a physical evaluation.However x-rays were provided.The x-rays were provided with a very poor quality and upon visual inspection of the x-rays, no evidence of implant fracture, disassociation, or anything indicative of a device nonconformance was found.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed, and the investigation may be re-opened as necessary.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE PS FB INSRT SZ 6 10MM
Common Device Name
ATTUNE IMPLANT : KNEE TIBIAL INSERT
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kara ditty-bovard
700 orthopaedic dr.
warsaw, IN 46581-0988
6107428552
MDR Report Key12705426
Report Number1818910-2021-23734
Device Sequence Number4550560
Product Code JWH
UDI-Device Identifier10603295050100
UDI-Public10603295050100
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K111433
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number1516-40-610
Device Catalogue Number151640610
Device Lot Number316244
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/14/2021
Supplement Date Received by Manufacturer12/09/2021
Initial Report FDA Received Date10/27/2021
Supplement Report FDA Received Date12/10/2021
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/10/2012
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ATTUNE FB TIB BASE SZ 6 CEM.; ATTUNE PS FB INSRT SZ 6 10MM.; ATTUNE PS FEM RT SZ 6 CEM.; COMPETITOR CEMENT.; ATTUNE FB TIB BASE SZ 6 CEM; ATTUNE PS FB INSRT SZ 6 10MM; ATTUNE PS FEM RT SZ 6 CEM; COMPETITOR CEMENT
Outcome Attributed to Adverse Event Required Intervention;
Patient Age57 YR
Patient SexMale
Patient Weight72 KG
-
-