| Model Number |
283889 |
| Medical Device Problem Code |
Device-Device Incompatibility (2919)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional narrative: udi: (b)(4).To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
|
| |
|
Event or Problem Description
|
|
It was reported by the customer in france that the barrel tornado burr 5.5mm device was defective; and that it could not snap onto the handpiece device.There was no procedure nor patient involvement reported.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: udi: (b)(4).To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.No additional information was provided.This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: investigation summary: according to the information provided, it was reported that the cutter cannot snap onto the micro tornado i shaver corn piece.Devices reported in this complaint were not received for evaluation, however; other devices reported under other customer complaints for the issue ¿cutter cannot snap onto the micro tornado shaver¿ were received and analyzed.A manufacturing record evaluation was performed for the finished device lot number: m2105003, and no non conformances were identified.The investigation was divided into two stages, 1st stage consisted in complaint burr visual inspection and functional test at the product analysis lab, the result for this test indicated that the devices could not be assembled into micro tornado handpiece (p/n 283512).Due to these results, the burrs were then measured, and it was identified that 2 burr features of the outer blade hub were over the upper limit.Devices were sent to r&d lab to continue with the investigation.The part underwent dimensional inspection, where it was confirmed that 2 burr features were oversized (12.575mm feature and 15.72mm feature).Manufacturing was notified about the condition and a nonconformance record was created to continue the investigation, root cause determination and mitigation of the issue.Depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Device history lot: a manufacturing record evaluation was performed for the finished device lot number: m2105003, and no non conformances were identified.
|
| |
|
Event or Problem Description
|
|
It was reported by the customer in france that during a shoulder arthroscopy procedure on an unknown date, it was observed that the barrel tornado burr 5.5mm device was defective; and that it could not snap onto the handpiece device.Another like device was used to complete the procedure with a delay of less than five minutes.There was no procedure nor patient involvement reported.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: b5: subsequent follow-up with the customer, additional information was received regarding the event.Therefore, the event description has been updated accordingly.
|
| |
|