MAUDE Adverse Event Report: ABBOTT VASCULAR NC TREK CORONARY DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR NC TREK CORONARY DILATATION CATHETER Back to Search Results
Model Number 1012445-20
Medical Device Problem Codes Material Separation (1562); Deformation Due to Compressive Stress (2889); Difficult to Advance (2920)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/07/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device is expected to be returned for investigation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.
 
Event or Problem Description
It was reported that the procedure was to treat a heavily calcified, heavily tortuous, 100% stenosed lesion in the proximal left circumflex (plcx) artery.The 2x20mm nc trek balloon dilatation catheter was advanced through the guide catheter (gc) with resistance met, and the bdc became kinked.While pulling the bdc back out of the guide catheter, the balloon separated from the delivery system and remained in the guide catheter.The guide catheter was removed with the separated delivery system as one unit.Another same size nc trek bdc was used to successfully dilatate the lesion without issue.There was no adverse patient effect and no clinically significant delay in the procedure.No additional information was provided.
 
Additional Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents from this lot.The investigation was unable to determine a conclusive cause for the reported difficulty advancing the device; however, the reported kink and separation appear to be related to operational context.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NC TREK CORONARY DILATATION CATHETER
Common Device Name
CORONARY DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR COSTA RICA, REG # 3009564766
52 calle 3 b31 coyol free zone
el coyol alajuela
CS  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key12706701
Report Number2024168-2021-09673
Device Sequence Number19831303
Product Code LOX
UDI-Device Identifier08717648151668
UDI-Public08717648151668
Combination Product (Y/N)N
Initial Reporter StateMS
Initial Reporter CountryUS
PMA/510(K) Number
K103153
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2024
Device Model Number1012445-20
Device Catalogue Number1012445-20
Device Lot Number10207G1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/07/2021
Supplement Date Received by Manufacturer12/28/2021
Initial Report FDA Received Date10/27/2021
Supplement Report FDA Received Date01/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/07/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-