• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, COSTA RICA LTDA PORTICO TAVI VALVE, 25 MM; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, COSTA RICA LTDA PORTICO TAVI VALVE, 25 MM; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED Back to Search Results
Model Number PRT-25
Device Problem Migration (4003)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/04/2021
Event Type  Injury  
Manufacturer Narrative
An event of migration was reported.A more comprehensive assessment could not be performed as the device remains implanted was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
It was reported that on (b)(6) 2021, a 25mm portico valve was selected for implant.During final release of the valve, the delivery system was turned very quickly, moving the implant position of the valve.The depth of the valve was reported to be 10mm.Therefore, a second 25mm portico valve was successfully implanted.The patient was reported to be in stable condition.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTICO TAVI VALVE, 25 MM
Type of Device
AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED
Manufacturer (Section D)
ST. JUDE MEDICAL, COSTA RICA LTDA
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 01897 -405
CS  01897-4050
Manufacturer (Section G)
ST. JUDE MEDICAL, COSTA RICA LTDA
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 01897 -405
CS   01897-4050
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key12706760
MDR Text Key278645030
Report Number3014918977-2021-00078
Device Sequence Number1
Product Code NPT
UDI-Device Identifier05415067012487
UDI-Public05415067012487
Combination Product (y/n)N
Reporter Country CodeAR
PMA/PMN Number
G120263
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 10/27/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/27/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date03/25/2022
Device Model NumberPRT-25
Device Catalogue NumberPRT-25
Device Lot Number7484176
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received10/04/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/25/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age77 YR
-
-