Related to 531449, 531880, 532090 and 532092 (same patient).The hospital reported that during a coronary artery bypass procedure, hst iii system (3.8mm) would not deploy properly.Finished case with another heartstring.No patient effects reported.
|
Trackwise # (b)(4).Updated sections: g4, g7, h2, h3, h6, h10, h11.Corrected sections: h3 - device evaluation changed from "device not returned" to "device discarded".Analysis of production: (3331/213/67) the device history records review concluded that there were no ncmrs, rework, or deviations documented for the reported lot number.Based on the dhr/lhr review results, it was determined that there is no relation between the manufacturing process and the reported failure.Historical data analysis: (4109/213/67) the review of the historical data indicates that no other similar complaints was reported for the same lot number and reported failure mode.Trend analysis: (4110/213/67) the overall 24 month product complaint trend data for the period nov-2019 through oct-2021 was reviewed.There were no triggers identified for the review period.Communication/interviews: (4111/213/67) communication/interviews were performed to obtain all possible information.Device discarded: (4115/3221/67) despite request and/ or customer indicated that the device was discarded; however, the actual device involved in the adverse event had been already discarded and thus irretrievably lost for testing.H3 other text : device discarded.
|