MAUDE Adverse Event Report: SMITH & NEPHEW, INC. LGN HK FEM ASSEMBLY SZ 7 LT; PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. LGN HK FEM ASSEMBLY SZ 7 LT; PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Model Number 71421377
Medical Device Problem Code Break (1069)
Health Effect - Clinical Codes Fall (1848); Failure of Implant (1924)
Date of Event 10/07/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference (b)(4).Postal code: (b)(4).
 
Event or Problem Description
It was reported that, after a tka surgery had been performed on (b)(6) 2020, the patient fell and the link mechanism of a lgn hk fem assembly sz 7 lt broke.This adverse event was solved by a revision surgery on (b)(6) 2021, in which the lgn hk fem assembly sz 7 lt was explanted and replaced.Current health status of patient is unknown.
 
Additional Manufacturer Narrative
The device was not returned for evaluation but the pictures were reviewed, and the failure mode was confirmed.The contribution of the device to the reported event could not be corroborated.The clinical/medical investigation concluded that based on the limited information provided, the root cause of the reported hinge knee component breakage is most likely ¿the patient fell and the link mechanism of a lgn hk fem assembly sz 7 lt broke.¿ the patient impact was the reported ¿patient fell,¿ device breakage and subsequent revision.Further patient impact beyond the reported events cannot be determined at this time and no further clinical medical assessment is warranted.A review of complaint history did not reveal similar events for the listed device.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.The instructions for use documents revealed this failure mode was previously identified.An historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.A potential causes for this event could include, but is not limited to, traumatic injury.Based on this investigation, the need for corrective action is not indicated.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LGN HK FEM ASSEMBLY SZ 7 LT
Common Device Name
PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key12708953
Report Number1020279-2021-07704
Device Sequence Number17283211
Product Code KRO
UDI-Device Identifier00885556414880
UDI-Public00885556414880
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K081111
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2021
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/13/2026
Device Model Number71421377
Device Catalogue Number71421377
Device Lot Number19BM11424
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/07/2021
Supplement Date Received by Manufacturer01/18/2022
Initial Report FDA Received Date10/28/2021
Supplement Report FDA Received Date01/21/2022
Date Device Manufactured03/13/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age70 YR
Patient SexMale
Patient Weight124 KG
-
-