SMITH & NEPHEW, INC. LGN HK FEM ASSEMBLY SZ 7 LT; PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER
|
Back to Search Results |
|
| Model Number |
71421377 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Codes |
Fall (1848); Failure of Implant (1924)
|
| Date of Event |
10/07/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Internal complaint reference (b)(4).Postal code: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that, after a tka surgery had been performed on (b)(6) 2020, the patient fell and the link mechanism of a lgn hk fem assembly sz 7 lt broke.This adverse event was solved by a revision surgery on (b)(6) 2021, in which the lgn hk fem assembly sz 7 lt was explanted and replaced.Current health status of patient is unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for evaluation but the pictures were reviewed, and the failure mode was confirmed.The contribution of the device to the reported event could not be corroborated.The clinical/medical investigation concluded that based on the limited information provided, the root cause of the reported hinge knee component breakage is most likely ¿the patient fell and the link mechanism of a lgn hk fem assembly sz 7 lt broke.¿ the patient impact was the reported ¿patient fell,¿ device breakage and subsequent revision.Further patient impact beyond the reported events cannot be determined at this time and no further clinical medical assessment is warranted.A review of complaint history did not reveal similar events for the listed device.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.The instructions for use documents revealed this failure mode was previously identified.An historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.A potential causes for this event could include, but is not limited to, traumatic injury.Based on this investigation, the need for corrective action is not indicated.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
| |
|
Search Alerts/Recalls
|
|
|