| Model Number |
305536 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/29/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Date of event: unknown.The date received by manufacturer has been used for this field.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported when using the bd precision glide¿ allergy syringe/needle the tip/luer of the syringe broke off.The following information was provided by the initial reporter.The customer stated: the medical professional reported, "the heads of the syringe are falling/breaking off.".
|
| |
|
Additional Manufacturer Narrative
|
|
H6: investigation summary: customer returned a photo of a 1/2ml syringe.Customer states that the heads of the syringe are falling/breaking off.The photo was examined and exhibited the hub-needle/shield assembly separated from the barrel.A review of the device history record was completed for batch # 0104220.All inspections were performed per the applicable operations qc specifications.There were zero (0) notifications noted that pertained to the complaint.Bd was able to duplicate or confirm the customer¿s indicated failure (hub separates).Capa (b)(4) has been opened to address this issue.H3 other text: see h10.
|
| |
|
Event or Problem Description
|
|
It was reported when using the bd precision glide¿ allergy syringe/needle the tip/luer of the syringe broke off.The following information was provided by the initial reporter.The customer stated: the medical professional reported, "the heads of the syringe are falling/breaking off.".
|
| |
|