|
Retainer ring = black the crest downloads were successful.Pump received with constant critical pump error (open book) and unsuccessful to download to thus.Unable to perform the functional test, including the self test, sleep current measurement, active current measurement, rewind test, prime/seating test, basic occlusion test, occlusion test, force sensor test and displacement test due to constant critical pump error (open book).Pump was cut/open to perform visual inspection and found moisture damage at the pcb1 and pcb2.Also moisture damage at the force sensor.Pump error 35 alarm found in the pump history due to force sensor moisture damage.On (b)(6) 2021 23:54:49.000 alarmalertnotification faultnumber = broken force sensor error (35).On (b)(6) 2021 00:04:00.000 alarmalertnotification faultnumber = broken force sensor error (35).Formatted history file lists data from (b)(6) 2021 to (b)(6) "202." there are no date listed on (b)(6) 2019.Unable to verify unexpected battery power loss anomaly or low battery alarm due to no date available on (b)(6) 2019.The long trace, pump history and sensor history was uploaded properly.However, the detail trace history unable to uploaded due to constant vibration and black display screen due moisture damage at the electronic assembly.Attempted to re - download 2nd time on thus by switching board.Original pcb2 and test pcb1.Try to download and unsuccessful to download to thus due moisture damage.Pump received with cracked select button keypad overlay.Test reservoir/pcap lock properly inside the reservoir tube.(b)(4).
|