• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MPRI SPRINT QUATTRO MRI SURESCAN; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MPRI SPRINT QUATTRO MRI SURESCAN; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT) Back to Search Results
Model Number 6946M55
Medical Device Problem Code Over-Sensing (1438)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/06/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that stored electrograms on a remote transmission showed possible intermittent t-wave oversensing (twos) on the right ventricular (rv) lead.The lead remains in use.No patient complications have been reported as a result of this event.
 
Event or Problem Description
Follow up with the patient's healthcare provider yielded that the lead was reprogrammed.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPRINT QUATTRO MRI SURESCAN
Common Device Name
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer (Section G)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key12711180
Report Number2649622-2021-21479
Device Sequence Number4921004
Product Code LWS
UDI-Device Identifier00643169335561
UDI-Public00643169335561
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
P920015
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/11/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date08/18/2018
Device Model Number6946M55
Device Catalogue Number6946M55
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/06/2021
Supplement Date Received by Manufacturer11/03/2021
Initial Report FDA Received Date10/28/2021
Supplement Report FDA Received Date11/11/2021
Date Device Manufactured10/06/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DDMB1D4 ICD, 5076-45; DDMB1D4 ICD, 5076-45
Outcome Attributed to Adverse Event Required Intervention;
Patient Age67 YR
Patient SexMale
Patient Weight53 KG
-
-