DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 220V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-00-3023-55 |
Device Problem
Device Displays Incorrect Message (2591)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Testing of actual/suspected device : a getinge field service engineer (fse) evaluated the iabp unit and did observe the leak in iab circuit alarm on the logs, with further diagnosis verified that there was also a safety disk malfunction.To fix the issue the fse replaced the safety disk from the customers purchased stock.He then performed all functional and safety checks to meet factory specifications.Unit passed all functional and safety test per factory specifications.The iabp was then released to the customer and cleared for clinical service.A supplemental report will be submitted upon completion of our investigation.The full event site name in is (b)(6).
|
|
Event Description
|
It was reported that during a routine check done by the customer the cs300 intra-aortic balloon pump (iabp) was giving a leak in iab circuit alarm.There was no patient involvement, and no adverse event reported.
|
|
Event Description
|
N/a.
|
|
Manufacturer Narrative
|
Updated sections: b4, g3, g6, g7, h2, h10, h11.Corrected sections: d1, d4, d5, h6(health clinical & impact).
|
|
Event Description
|
N/a.
|
|
Manufacturer Narrative
|
Updated fields: b4, g3, g6, h2, h4, h6, h10.
|
|