• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. ULTRAVERSE 035; PTA BALLOON DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. ULTRAVERSE 035; PTA BALLOON DILATATION CATHETER Back to Search Results
Model Number U357584
Medical Device Problem Codes Material Rupture (1546); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/29/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As the lot number for the device was provided, a review of the device history record is currently being performed.The device has not been returned to the manufacturer for evaluation.However, a photo was provided for review.The investigation of the reported event is currently underway.(expiry date: 01/2024).Device not returned.
 
Event or Problem Description
It was reported that during an angioplasty procedure, the balloon allegedly ruptured and detached.Further reported that the physician removed the detached piece of balloon.There was no reported patient injury.
 
Event or Problem Description
It was reported that during an angioplasty procedure, the balloon allegedly ruptured at 18 atm.It was further reported that the balloon allegedly detached and the detached piece of balloon was removed.There was no reported patient injury.
 
Additional Manufacturer Narrative
H10: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: the sample was not returned to the manufacturer for evaluation.However, one photo was reviewed.The photo shows the detached part of the catheter along with the portion of the ruptured balloon (compound rupture) attached to it.The ruptured balloon appears bloody.No other specific anomalies noted.Therefore based on the submitted photo, compound balloon rupture and balloon detachment can be confirmed since the detached catheter and ruptured balloon can be seen.Therefore the investigation can be confirmed for the reported balloon rupture since a compound rupture was noted on the balloon.The investigation can be confirmed for the reported detachment since the device is noted to be detached distally.A definitive root cause for the alleged compound balloon rupture and detachment could not be determined based upon the provided information.A definitive root cause could not be determined based upon the available information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: b5, d4 (expiry date: 01/2024), g3, h6 (method).H11: h6 (result, conclusion).H11: section a through f ¿ the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
 
Additional Manufacturer Narrative
H10: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: one photo was reviewed.The photo shows the detached distal part of the catheter exposing the inner guide wire lumen along with the portion of the ruptured balloon (circumferential rupture) attached to it.The ruptured balloon appears bloody.No other specific anomalies noted.Therefore based on the submitted photo, balloon rupture and balloon detachment can be confirmed since the detached catheter and ruptured balloon can be seen.One ultraverse035 pta dilatation catheter in two segments has returned for evaluation.On the visual evaluation of the device the segment 1 consist detached distal end of the catheter exposing the inner guide wire lumen and circumferentially ruptured balloon.Segment 2 consist of the remaining part of the catheter.Only the proximal marker band was present.All the anomalies noted on the microscopic observation.No other specific anomalies noted.No functional testing performed due to the condition device.Therefore, the investigation was confirmed for the reported balloon rupture as the returned device has circumferential rupture.The investigation was also confirmed reported detachment as the catheter retuned in 2 segments for evaluation.A definitive root cause for the alleged circumferential balloon rupture and detachment could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: b5, d4 (expiry date: 01/2024).H11: section a through f ¿ the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that during an angioplasty procedure, the pta balloon allegedly ruptured at 18 atm.It was further reported that the balloon allegedly detached.Reportedly, another balloon was inflated to remove the detached piece through the introducer.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRAVERSE 035
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
CLEARSTREAM TECHNOLOGIES LTD.
moyne upper
enniscorthy, co. wexford N A
EI   N A
Manufacturer Contact
judy ludwig
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key12711482
Report Number2020394-2021-01919
Device Sequence Number17337052
Product Code LIT
UDI-Device Identifier00801741093951
UDI-Public(01)00801741093951
Combination Product (Y/N)N
Initial Reporter CountryBE
PMA/510(K) Number
K142261
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberU357584
Device Catalogue NumberU357584
Device Lot NumberCMFP0198
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/22/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 09/30/2021
Supplement Date Received by Manufacturer11/17/2021
12/20/2021
Initial Report FDA Received Date10/28/2021
Supplement Report FDA Received Date11/19/2021
12/21/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-