• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - FAIRFIELD SENSATION PLUS 8FR. 50CC IAB; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - FAIRFIELD SENSATION PLUS 8FR. 50CC IAB; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0684-00-0576-01
Device Problem Optical Problem (3001)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/08/2021
Event Type  malfunction  
Manufacturer Narrative
The device will not be returned to the manufacturer so we are unable to complete an evaluation.If provided we will send a supplemental report with our additional findings.(b)(4).
 
Event Description
It was reported that after inserting the intra-aortic balloon (iab), the console immediately indicated that there was a fiber optic sensor failure.Troubleshooting was unsuccessful at relieving the sensor failure message.The customer was then advised to transduce the central lumen.This provided a clean, crisp waveform and therapy was continued.The patient was later transferred to another facility.There was no patient harm or adverse event reported.
 
Manufacturer Narrative
The device has not been returned to the manufacturer so we are unable to complete an evaluation.(b)(4).Reference complaint #(b)(4).
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENSATION PLUS 8FR. 50CC IAB
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - FAIRFIELD
15 law drive
fairfield NJ
Manufacturer (Section G)
DATASCOPE CORP. - FAIRFIELD
15 law drive
fairfield NJ
Manufacturer Contact
brian schaeffer
15 law drive
fairfield, NJ 
MDR Report Key12711839
MDR Text Key278789044
Report Number2248146-2021-00708
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108605
UDI-Public10607567108605
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112327
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 12/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/28/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0684-00-0576-01
Device Catalogue Number0684-00-0575
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/02/2021
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Treatment
UNKNOWN.; UNKNOWN.
Patient Age60 YR
Patient SexMale
Patient Weight79 KG
-
-