• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC 4.5 VA-LCP CURVED COND PL/8 HOLE/195/RT; CONDYLAR PLATE FIXATION IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC 4.5 VA-LCP CURVED COND PL/8 HOLE/195/RT; CONDYLAR PLATE FIXATION IMPLANT Back to Search Results
Model Number 02.124.408
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Bone Fracture(s) (1870); Unspecified Infection (1930)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional product code: hrs, hwc.Complainant part is not expected to be returned for manufacturer review/investigation.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that, a synthes va(variable angle) distal femur plate and screws will be removed.The fracture re-reduced and revised on (b)(6) 2021.The reason is a failed peri prosthetic fracture and the hardware was removed completely.This report is for (1) 4.5 va-lcp curved cond pl/8 hole/195/rt.This is report 1 of 9 for complaint (b)(4).
 
Additional Manufacturer Narrative
Product complaint (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Investigation summary: the complaint device was not received for investigation.A photo investigation was performed based on the x-ray image received.The x-ray image was reviewed, and the complaint condition is not confirmed.The implanted plate shows no signs of breakage or deformation in the provided image.A definitive assignable root cause could not be determined based on the provided information.As the device was not returned, an as-received condition could not be assessed and a dimensional inspection and document/specification review were not completed.No lot number was provided or found in the photos, therefore no dhr could be performed.If a valid lot number is provided or the physical device returned, the dhr will be revisited.During the investigation, no product design issues or discrepancies were observed (based on the image) that may have contributed to the complaint condition.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Device history lot : no lot number was provided or found in the photos, therefore no dhr could be performed.If a valid lot number is provided or the physical device returned, the dhr will be revisited.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
4.5 VA-LCP CURVED COND PL/8 HOLE/195/RT
Common Device Name
CONDYLAR PLATE FIXATION IMPLANT
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key12712072
Report Number2939274-2021-06344
Device Sequence Number4560502
Product Code JDP
UDI-Device Identifier10886982042774
UDI-Public(01)10886982042774
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K083025
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/03/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number02.124.408
Device Catalogue Number02.124.408
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 10/03/2021
Supplement Date Received by Manufacturer11/01/2021
Initial Report FDA Received Date10/28/2021
Supplement Report FDA Received Date11/02/2021
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNK - SCREWS: 4.5 MM CORTEX; UNK - SCREWS: 4.5 MM CORTEX; UNK - SCREWS: 4.5 MM CORTEX; UNK - SCREWS: 4.5 MM CORTEX; UNK - SCREWS: 5.0 MM CANNULATED CONICAL; UNK - SCREWS: 5.0 MM VA LOCKING; UNK - SCREWS: 5.0 MM VA LOCKING; UNK - SCREWS: 5.0 MM VA LOCKING; UNK - SCREWS: 4.5 MM CORTEX; UNK - SCREWS: 4.5 MM CORTEX; UNK - SCREWS: 4.5 MM CORTEX; UNK - SCREWS: 4.5 MM CORTEX; UNK - SCREWS: 5.0 MM CANNULATED CONICAL; UNK - SCREWS: 5.0 MM VA LOCKING; UNK - SCREWS: 5.0 MM VA LOCKING; UNK - SCREWS: 5.0 MM VA LOCKING
Outcome Attributed to Adverse Event Required Intervention;
Patient Weight102
-
-