ONKOS SURGICAL ELEOS; SEGMENTAL STEM, CEMENTED, 11X120MM, STRAIGHT, FLUTED, 28MM COLLAR Back to Search Results
Model Number CS-11120-03M
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Unspecified Infection (1930); Fluid Discharge (2686)
Date of Event 10/04/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The investigation is in process.The device will not be returned for evaluation.When the investigation is complete, a supplemental mdr will be submitted accordingly.Multiple mdrs were submitted for this event: #3013450937-2021-00237.#3013450937-2021-00238.#3013450937-2021-00240.#3013450937-2021-00241.#3013450937-2021-00242.#3013450937-2021-00243.#3013450937-2021-00244.#3013450937-2021-00245.
 
Event or Problem Description
It was reported that the patient underwent a revision surgery on (b)(6) 2021 due to an alleged infection.The patient's knee was discharging fluid and the surgeon wanted to test for an infection.During the revision surgery, the surgeon performed a poly swap and a washout.The following eleos implants were revised during the revision surgery: tibial hinge component, distal femur axial pin, and tibial poly spacer.The following eleos implants remain implanted from the patient's original surgery: distal femur, male-female midsection, cemented segmental stem, tibial baseplate, cemented stem extension, and resurfacing patella.No additional information regarding this event has been provided.
 
Additional Manufacturer Narrative
This report is being submitted to include additional information.The investigation is complete.The device was not returned for evaluation.The root cause of the alleged infection could not be determined.The device history records and sterilization batch records were reviewed and no issues during manufacturing or sterilization were identified that would have contributed to this complaint.If any additional information is obtained, a supplemental mdr will be filed accordingly.The following sections were updated: b4: date of this report added.G3: date received by manufacturer added.G6: type of report added.H2: follow-up type added.H3: device evaluated by manufacturer updated to no.H6: type of investigation code updated to 3331: analysis of production records.H6: type of investigation code updated to 4111: communication/interviews.H6: type of investigation code updated to 4110: trend analysis.H6: type of investigation code updated to 4114: device not returned.H6: type of investigation code updated to 4117: device not accessible for testing.H6: type of investigation code updated to 4109: historical data analysis.H6: investigation findings code updated to 213: no device problem found.H6: investigation conclusions code updated to 4315: cause not established.H10: additional narratives/data.
 
Event or Problem Description
It was reported that the patient underwent a revision surgery on (b)(6) 2021 due to an alleged infection.The patient's knee was discharging fluid and the surgeon wanted to test for an infection.During the revision surgery, the surgeon performed a poly swap and a washout.The following eleos implants were revised during the revision surgery: tibial hinge component, distal femur axial pin, and tibial poly spacer.The following eleos implants remain implanted from the patient's original surgery: distal femur, male-female midsection, cemented segmental stem, tibial baseplate, cemented stem extension, and resurfacing patella.No additional information regarding this event has been provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELEOS
Common Device Name
SEGMENTAL STEM, CEMENTED, 11X120MM, STRAIGHT, FLUTED, 28MM COLLAR
Manufacturer (Section D)
ONKOS SURGICAL
77 east halsey road
parsippany NJ 07054
Manufacturer (Section G)
PHILLIPS PRECISION INC
7 paul kohner pl
elmwood park NJ 07407
Manufacturer Contact
sara dailey
77 east halsey road
parsippany, NJ 07054
MDR Report Key12712515
Report Number3013450937-2021-00239
Device Sequence Number20079911
Product Code KRO
UDI-Device IdentifierB278CS1112003M0
UDI-Public+B278CS1112003M0
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K161520
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberCS-11120-03M
Device Catalogue NumberCS-11120-03M
Device Lot Number88302
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/29/2021
Supplement Date Received by Manufacturer12/21/2021
Initial Report FDA Received Date10/28/2021
Supplement Report FDA Received Date12/21/2021
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/11/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
P/N 25000009E, ELEOS DISTAL FEMUR; P/N 25001040E, ELEOS MALE-FEMALE MIDSECTION; P/N 25001208E. ELEOS TIBIAL POLY SPACER; P/N 25002111E, ELEOS DISTAL FEMUR AXIAL PIN; P/N 25002202E, ELEOS TIBIAL BASEPLATE; P/N KPONTP32E, ELEOS RESURFACING PATELLA; P/N KSC01530E, ELEOS CEMENTED STEM EXTENSION; P/N THSMWOS01M, ELEOS TIBIAL HINGE COMPONENT; P/N 25000009E, ELEOS DISTAL FEMUR; P/N 25001040E, ELEOS MALE-FEMALE MIDSECTION; P/N 25001208E. ELEOS TIBIAL POLY SPACER; P/N 25002111E, ELEOS DISTAL FEMUR AXIAL PIN; P/N 25002202E, ELEOS TIBIAL BASEPLATE; P/N KPONTP32E, ELEOS RESURFACING PATELLA; P/N KSC01530E, ELEOS CEMENTED STEM EXTENSION; P/N THSMWOS01M, ELEOS TIBIAL HINGE COMPONENT
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age87 YR
Patient SexFemale
-
-