| Model Number |
442193 |
| Medical Device Problem Code |
False Positive Result (1227)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/29/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A device evaluation and/or device history review is anticipated, but is not complete.Upon completion, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that while using bd bactec¿ plus anaerobic/f culture vials 6 false positive results were obtained by the laboratory personnel.A gram stain was used to confirm the results as false positives.There was no indication that results were reported out and there was no report of patient impact.The following information was provided by the initial reporter: "it was reported that there are false positives.".
|
| |
|
Event or Problem Description
|
|
It was reported that while using bd bactec¿ plus anaerobic/f culture vials 6 false positive results were obtained by the laboratory personnel.A gram stain was used to confirm the results as false positives.There was no indication that results were reported out and there was no report of patient impact.The following information was provided by the initial reporter: "it was reported that there are false positives.".
|
| |
|
Additional Manufacturer Narrative
|
|
The following fields were updated due to additional information: d.9.Device available for eval?: yes.D.9.Returned to manufacturer on: 10/15/2021.H.6.Investigation: customer reported a false positive defect.Bd was unable to reproduce customer experience with the bactec product.A false positive response was not observed when retention and returned good samples were tested.Batch history records were reviewed, and all testing were within specification for product release.Users are cautioned in the package insert under limitation of the procedure: ¿a gram-stained smear from culture medium may contain small number of non-viable organisms derived from media constituents, staining reagents, immersion oil, glass slide and specimens used for inoculation.There are many factors that can influence the false positive rate, including blood volume, blood cell counts, environmental factors, and media lot to lot variations.Complaint is unconfirmed based on retention and returned good samples results and batch history record review.No correctives actions were required.A cross functional team continually monitors all product complaints for trends and determines if any additional actions are necessary beyond the current investigation process.H3 other text : see h.10.
|
| |
|
Search Alerts/Recalls
|