|
The pacemaker was received for analysis on 12-jul-2021.Prior to the analysis of the device, the quality documents accompanying the manufacturing process for this pacemaker were re-investigated and all production steps were performed accordingly.Particularly the final acceptance test proved the device functions to be as specified.The device interrogation revealed the eri battery status, confirming the clinical observation, which was triggered on (b)(6) 2020.The pacemaker was implanted for approximately 120 months.The amount of charge taken from the battery was verified and the battery condition was found to be normal and as expected based on the programmed settings and the overall current consumption.The ability of the device to deliver therapies was verified.The anti-bradycardia pacing pulses in eri mode proved to be normal and in amplitude and frequency as programmed.There was no indication of a device malfunction.Further analysis showed that the pacemaker triggered the eri battery status correctly based on the remaining battery capacity and the programmed parameters.However, the displayed time to eri during follow-up was not consistent with the actual eri detection of this pacemaker.The root cause for this observation could not be conclusively determined.Please note, that in general, a minimum of 6 months remaining service time starting from eri detection is guaranteed, independent of the programmed parameters.Thus, a specific proportion of battery capacity has to be reserved by the pacemaker for the service time after eri detection.Due to a conservative battery management an additional 6 months safety margin is provided, resulting in a typical remaining service time of about 12 months after eri detection.In conclusion, despite the inconsistency of the time to eri display during follow-up and the actual eri detection by the pacemaker, there was no risk for the patient, as there would have been sufficient remaining service time left after eri detection.The root cause for this observation, however, is not conclusively determined.
|