Model Number M00565060
Medical Device Problem Codes Premature Activation (1484); Device Misassembled During Manufacturing /Shipping (2912)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/06/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).The complainant indicated that the device was disposed and will not be returned for evaluation; therefore, a problem analysis of the complaint device could not be completed.If any further relevant information is identified, a supplemental medwatch will be filed.
 
Event or Problem Description
It was reported to boston scientific corporation on (b)(6) 2021 that a wallflex enteral colonic stent was to be implanted to treat a malignant short intrinsic colonic stricture in the distal sigmoid colon during a stent placement procedure performed on (b)(6) 2021.Reportedly, the patient's anatomy was tortuous.During the procedure, upon deploying the stent, the stent prematurely deployed and was placed beyond the stricture.Additionally, it was noted that the delivery system did not have a white marker to indicate the point of no return.A colectomy surgery was performed to resect part of the patient's colon where the stent was deployed and to remove the stent.The patient was admitted to the hospital beyond standard of care.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WALLFLEX COLONIC
Common Device Name
STENT, COLONIC, METALIC, EXPANDABLE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
MDR Report Key12712870
Report Number3005099803-2021-05591
Device Sequence Number17319421
Product Code MQR
UDI-Device Identifier08714729456537
UDI-Public08714729456537
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K061877
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 10/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/05/2022
Device Model NumberM00565060
Device Catalogue Number6506
Device Lot Number0026125936
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/08/2021
Initial Report FDA Received Date10/28/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-