• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER INFINITI VISION SYSTEM, ACCESSORY, HANDPIECE; UNIT, PHACOFRAGMENTATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER INFINITI VISION SYSTEM, ACCESSORY, HANDPIECE; UNIT, PHACOFRAGMENTATION Back to Search Results
Catalog Number 8065750469
Device Problem Suction Failure (4039)
Patient Problems Unspecified Infection (1930); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 10/01/2021
Event Type  Injury  
Manufacturer Narrative
This is the first of two reports for this reported event.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A nurse reported that during surgery the system presented with poor aspiration in phacoemulsification mode they changed the handpiece out twice and on the third one the aspiration functioned as expected.The customer also noted hot/warm handpieces.The procedure was completed.Additional information was received indicating that the hot/warm hp's were not allowed to cool after coming from the autoclave.The patient presented to the emergency room on post operative day two.The patient was diagnosed with periorbital cellulitis in the left eye.The patient required administration of intravenous antibiotics and antibiotics by mouth.
 
Manufacturer Narrative
The company service representative examined the system, and no problems were found with the system.The system was then tested and met all product specifications.The company service representative in-serviced the staff on setting up the system for cases along with basic troubleshooting.The company service representative focused on cooling times of the phaco handpieces and making sure phaco handpieces are at room temperature before using.The phaco handpiece was not returned for evaluation.Specific product identifiers (lot number, batch number, and/or serial number) were not provided and could not be determined at this time.However, all device history records are reviewed prior to product release to ensure the product was manufactured in compliance with the device master record and meets release criteria.A review for complaints reported against this lot/batch/serial number cannot be performed as the lot/batch/serial number is unknown.The product under investigation is not a serviceable device.Therefore, a service record review was not performed for the handpieces.The root cause of the reported event is inconclusive.Manufacturer will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFINITI VISION SYSTEM, ACCESSORY, HANDPIECE
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer (Section G)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key12714215
MDR Text Key278873697
Report Number2028159-2021-01240
Device Sequence Number1
Product Code HQC
UDI-Device Identifier00380657504695
UDI-Public00380657504695
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112425
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 03/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/28/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number8065750469
Device Lot NumberASKU
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/18/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
INFINITI VISION SYSTEM; INFINITI VISION SYSTEM
Patient Outcome(s) Required Intervention; Other;
-
-