| Model Number |
561CA |
| Medical Device Problem Code |
Degraded (1153)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/19/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The manufacturer received information alleging an issue related to a cpap device's sound abatement foam.There was no report of patient harm or injury this issue was reported to the fda per 21 cfr 806.The device will be corrected per res 88058.
|
| |
|
Additional Manufacturer Narrative
|
|
The manufacturer was contacted in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.The manufacturer previously received information alleging visualization of particles related to a cpap device's sound abatement foam.There was no report of patient harm or injury.After the third attempt to have the device and components returned for evaluation, the customer declined to respond to the gfe related questions and terminated the call.The manufacturer is submitting a follow-up report at this time.If pertinent information becomes available to the manufacturer at a later date, an addendum to this follow-up report will be filed.
|
| |
|
Additional Manufacturer Narrative
|
|
The remstar auto a-flex device was returned to a third-party service center for further evaluation.During evaluation, the device was found to be functional.Sound abatement foam degradation was not observed.There was no evidence of foam degradation or particles found in the assembly.The device did not contain any type of contamination the service center also retrieved and reviewed the device's error logs.There were 7 errors found.The third-party service center could not confirm the customer's allegation and there was no visible foam degradation.The unit was scrapped after device evaluation was completed.Sections g1, h2, h3, and h6 have been updated in this report.
|
| |
|
Search Alerts/Recalls
|