• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V. BD Q-SYTE LUER ACCESS SPLIT SEPTUM; INTRAVASCULAR ADMINISTRATION SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V. BD Q-SYTE LUER ACCESS SPLIT SEPTUM; INTRAVASCULAR ADMINISTRATION SET Back to Search Results
Catalog Number 395242
Device Problem Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/29/2021
Event Type  malfunction  
Manufacturer Narrative
Medical device expiration date: unknown.Device manufacture date: unknown.Investigation summary: bd was unable to perform a thorough investigation as no sample, material number, lot, or batch number were provided.Examination of the product involved may provide clarification as to the cause for the reported failure.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored monthly.Our business team regularly reviews the collected data for identification of emerging trends.
 
Event Description
It was reported bd q-syte luer access split septum detached from the stopcock.The following information was provided by the initial reporter, translated from (b)(6): "when terumo's syringe 20 ml was connected to the stopcock, the connection was loose and the syringe rotated.When pulled, the syringe was detached from the stopcock.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD Q-SYTE LUER ACCESS SPLIT SEPTUM
Type of Device
INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V.
periferico luis donaldo
colosio no. 579
nogales
MX 
MDR Report Key12715763
MDR Text Key279011867
Report Number9610847-2021-00522
Device Sequence Number1
Product Code FPA
Combination Product (y/n)N
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,other,user facility
Type of Report Initial
Report Date 10/07/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number395242
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 09/29/2021
Initial Date FDA Received10/28/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-