It was reported, that during a tka a genesis ii oval resurf pat trial 35mm broke inside the patient.All pieces were retrieved.Surgery was completed, without delay, by using a back-up device.The patient was not harmed beyond what has been described.
|
|
H3, h6: the device was not returned for evaluation and the reported event could not be confirmed.The contribution of the device to the reported event could not be corroborated.The clinical/medical investigation concluded that, a genesis ii oval resurf pat trial 35mm broke and all pieces were retrieved.No clinically relevant supporting documentation was provided for review.Based on the limited information provided, the root cause of the breakage could not be determined.It was reported that the procedure was completed without surgical delay, using a back-up device.No patient harm or complications were reported.No further medical assessment can be rendered at this time.A review of complaint history revealed similar events for the listed device, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.
|