|
It was reported that the patient underwent an initial lumber fixation procedure with the assistance of navigation.During the procedure, following placement of the pedicle screws, the surgeon had concerns with the navigation accuracy.The screw placement was reviewed under fluoroscopy and two (2) misplaced screws were observed.The surgeon repositioned the screws using fluoroscopy to complete the procedure.No patient impact was reported.
|
|
The device was evaluated on site and during evaluation, the reported problem could not be confirmed or replicated, and no other device component or functional issues were found.Therefore, the cause of the reported event is unknown.However, this problem type is currently being investigated through the capa process.Labeling review: "warnings: 1.Read all instructions and understand all warnings and cautions before using the system or accessories.Failure to do so may lead to serious medical consequences.Refer to the instructions for use accompanying other nuvasive products before use with the system to assure proper and safe use of these devices.3.If system navigation seems inaccurate and re-registration steps to restore accuracy are not successful, abort use of the system.4.Assess navigational accuracy repeatedly throughout a procedure when using a surgical navigation system.A.Reconfirm accuracy by positioning the navigated instrument tip on an identifiable anatomical landmark and comparing the actual tip location to that displayed by the system.B.If the stereotaxic navigation system does not appear to be accurate despite troubleshooting (e.G., resetting the system), do not rely on the navigation system.5.If data acquisition seems inaccurate, software application does not initiate or malfunctions during use, and recommended steps to restore the system are not successful ¿ abort use of the system.6.The pulse navigation system has been successfully tested against the requirements of iec 60601-1-2.However, radio frequency (rf) interference could hamper its operation or the operation of other nearby electrical devices.If you suspect either of these conditions, move the conflicting equipment farther apart, separate the equipment with a rf barrier, or discontinue use of the system.7.Movement of the patient during 3d radiographic image acquisition may result in inaccurate measurements.Efforts to immobilize the patient during data acquisition should be taken to restrict patient movement.If patient movement during data acquisition is observed or suspected, re-start the data acquisition process to ensure accurate image.8.Scans from only calibrated intraoperative 3d imaging systems must be used for pulse navigation.Use of scans from uncalibrated imaging systems may lead to navigation error.37.Capturing the patient reference arrays for registration can happen before you drape or after you remove the drape from the patient reference arrays during 3d image acquisition for navigation.Make sure to gently drape over the patient reference arrays, ensuring the arrays are not bumped or moved in this process.Any movement of the arrays during this process could require a re-capture of the arrays, or even re-scan of the patient if the registration accuracy is not adequate"."precautions: 14.The accuracy of data acquisition can be influenced by the rotation of the passive reflective array relative to the camera.Cameras should be placed such that a direct line of sight to the operative area is established"."maintenance: pulse is installed and qualified on site by a certified nuvasive technician.Prior to each use, the system should be inspected and tested as follows: pulse system inspect the touch screen for damage and confirm that all labels are legible and in good condition.Confirm the power cord insulation is in good condition and that there are no exposed wires.If any damage is observed, the device must be taken out of service and returned to nuvasive if the application freezes during use, restart the machine by going to the windows start menu and selecting restart".9511816 rev.E.
|