| Model Number |
H700169 |
| Medical Device Problem Code |
Communication or Transmission Problem (2896)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/27/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
During an atrial fibrillation procedure, signals could not be displayed.An error message was noted stating: check the amplifier and connections to the display workstation.Troubleshooting involved multiple restart of all devices, cables were checked, the converter was checked and the fiber optic cable was replaced which did not resolve the issue.The procedure was cancelled and the patient was taken to another room where carto was performed with no consequences to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
The results of the investigation are inconclusive as the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.No non-conformances associated with the reported event were identified.Based on the information received, the cause of the reported communication issue and subsequent cancellation remains unknown.
|
| |
|