Model Number H700169
Device Problem Communication or Transmission Problem (2896)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/27/2021
Event Type  malfunction  
Event Description
During an atrial fibrillation procedure, signals could not be displayed.An error message was noted stating: check the amplifier and connections to the display workstation.Troubleshooting involved multiple restart of all devices, cables were checked, the converter was checked and the fiber optic cable was replaced which did not resolve the issue.The procedure was cancelled and the patient was taken to another room where carto was performed with no consequences to the patient.
 
Manufacturer Narrative
The results of the investigation are inconclusive as the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.No non-conformances associated with the reported event were identified.Based on the information received, the cause of the reported communication issue and subsequent cancellation remains unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WORKMATE¿ CLARIS¿ SYSTEM AMPLIFIER
Type of Device
COMPUTER, DIAGNOSTIC, PROGRAMMABLE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
one st. jude medical drive
st. paul MN 55117
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.
one st. jude medical drive
st. paul MN 55117
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key12719272
MDR Text Key279014133
Report Number2184149-2021-00342
Device Sequence Number1
Product Code DQK
UDI-Device Identifier05415067001696
UDI-Public05415067001696
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K151911
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberH700169
Device Catalogue NumberH700169
Device Lot Number7843439
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/27/2021
Initial Date FDA Received10/29/2021
Supplement Dates Manufacturer Received12/29/2021
Supplement Dates FDA Received01/06/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/05/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-