SYNTHES GMBH HAMMER GUIDE F/TEN; HAMMER, SURGICAL Back to Search Results
Catalog Number 359.218
Medical Device Problem Code Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/20/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Complainant part is expected to be returned for manufacturer review/investigation but has yet to be received.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Device report from synthes reports an event in (b)(6) as follows: it was reported that on (b)(6) 2021 the customer reports knocked out thread.This report is for one (1) hammer guide f/ten.This is report 2 of 2 for complaint (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H3, h6: the device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Visual inspection: the complaint device (hammer guide f/ten) was returned for investigation.Examination of the returned device found damaged threads, confirming the reported condition.No other issues were identified.Functional test: the complaint device (hammer guide f/ten) cannot be threaded correctly to the inserter, remains loose.Document/specification review: dimensional inspection was performed for the design of device.Investigation conclusion: the reported condition of the complaint device (hammer guide f/ten) is confirmed.Examination of the returned device found damaged threads.There was no indication that a design or manufacturing issue contributed to the complaint.No design issues were observed during the document/ specification review.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Device history lot - a manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HAMMER GUIDE F/TEN
Common Device Name
HAMMER, SURGICAL
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
SYNTHES HAEGENDORF GMBH-CN
im bifang 6
haegendorf 4614
SZ   4614
Manufacturer Contact
kara ditty-bovard
1302 wright lane east
west chester, PA 19380
6107195000
MDR Report Key12719475
Report Number8030965-2021-09115
Device Sequence Number19173221
Product Code FZY
UDI-Device Identifier07611819104365
UDI-Public(01)07611819104365
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number359.218
Device Lot Number44P3658
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/19/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 10/20/2021
Supplement Date Received by Manufacturer11/19/2021
01/17/2022
Initial Report FDA Received Date10/29/2021
Supplement Report FDA Received Date12/17/2021
01/17/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/27/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
INSERTER F/TEN.; INSERTER F/TEN
Patient SexUnknown
-
-