| Model Number |
ML02-0117 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Skin Inflammation/ Irritation (4545)
|
| Date of Event |
10/15/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The duration of the reported skin redness and irritation was not provided.The technician reported that the cap of the bottle had been loose and that the box containing the bottle of rapicide pa part a had been wet, likely due to chemistry leaking prior to the event.Following the event, the user contacted (b)(6) and was provided the rapicide pa part a sds.No additional medical intervention was sought or required.It is not known if the user was wearing proper ppe at the time of the reported event.The rapicide pa part a sds provides the following warning language and guidance, related to skin exposure and ppe usage: "causes severe skin burns and eye damage.Wash hands thoroughly after handling.Wear protective gloves/protective clothing/eye protection/face protection.If on skin (or hair): rinse skin with water/shower." furthermore, rapicide pa part a is bottled with a vented lid to allow chemical fumigation.The following language is from the rapicide pa label, "store upright in shipping carton.Store in original closed container.Never tamper with vent." it is unknown if the chemistry bottle orientation was changed during transport or storage at the facility.No additional reports of adverse event or harm were received.The bottle subject of the event was not returned for evaluation.Medivators will continue to monitor for similar events to ensure the product continues to perform as expected.
|
| |
|
Event or Problem Description
|
|
A user facility reported that a technician removed a bottle of medivators rapicide pa part a from the box and was splashed by the chemistry on their neck and arm.The technician took a shower to rinse off the chemicals and noted slight redness and irritation of the skin on her arm and neck due to the chemistry exposure.It is unknown if ppe was worn at the time of the event.The technician reported that the chemistry bottle cap had been loose, resulting in the reported event.No additional reports of injury were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The duration of the reported skin redness and irritation was not provided.The technician reported that the cap of the bottle had been loose and that the box containing the bottle of rapicide pa part a had been wet, likely due to chemistry leaking prior to the event.It is not known if the user was wearing proper ppe at the time of the reported event.The rapicide pa part a sds provides the following warning language and guidance, related to skin exposure and ppe usage: "causes severe skin burns and eye damage.Wash hands thoroughly after handling.Wear protective gloves/protective clothing/eye protection/face protection.If on skin (or hair): rinse skin with water/shower." furthermore, rapicide pa part a is bottled with a vented lid to allow chemical fumigation.The following language is from the rapicide pa label, "store upright in shipping carton.Store in original closed container.Never tamper with vent." it is unknown if the chemistry bottle orientation was changed during transport or storage at the facility.No additional reports of adverse event or harm were received.The bottle subject of the event was not returned for evaluation.Medivators will continue to monitor for similar events to ensure the product continues to perform as expected.
|
| |
|
Search Alerts/Recalls
|