• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AESCULAP AG SILICONE RING WITH NEEDLE; HANDHELD PRODUCTS & LIGATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AESCULAP AG SILICONE RING WITH NEEDLE; HANDHELD PRODUCTS & LIGATION Back to Search Results
Model Number PL595SU
Medical Device Problem Codes Fracture (1260); Material Integrity Problem (2978)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/01/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Manufacturing site evaluation: investigation on-going.Should relevant additional information / investigation results become available, a supplemental medwatch report will be submitted.
 
Event or Problem Description
It was reported to aesculap ag tha a silicone ring with needle (part # pl595su) was used during a procedure performed on (b)(6) 2021.According to the complainant, it was reported that the silicone ring was torn.Reportedly, the customer held the needle and set it, but the silicone ring tore.The complaint device was returned to the manufacturer for evaluation.No patient complications were reported as a result of the event.Although requested, additional information has not been made available.The adverse event / malfunction is filed under aag reference (b)(4).Associated medwatch-reports: 9610612-2021-00683 (400530801 pl595su).9610612-2021-00684 (400530802 pl595su).
 
Additional Manufacturer Narrative
Associated medwatch reports: pl595su - 400530801 (9610612-2021-00683), pl595su - 400530802 (9610612-2021-00684).Investigation: visual investigation: the analysis of the damage pattern shows a fracture due to overload.No pores, inclusions or foreign bodies could be found at tht damage site.Batch history review: the device quality and manufacturing history records (dhr) have been checked for all leading device(s) lot numbers and the products found to be according to our specification valid at the time of production.There are 2 similar complaints against the same lot number(s).The review of risk assessment revealed that the overall risk level (severity 6(10) x probability of occurrence 3(10)) according to din en iso 14971 is still acceptable.Explanation and rationale: based on the damage pattern, the breakage of the silicone ring was most likely caused by overload by pulling beyond yield strenth.Conclusion and measures / preventive measures: based upon the investigation results a clear root cause conclusion cannot be drawn.There is no indication for a material defect or manufacturing failure on the basis of the device history records.Based on the investigation results, a capa is not necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SILICONE RING WITH NEEDLE
Common Device Name
HANDHELD PRODUCTS & LIGATION
Manufacturer (Section D)
AESCULAP AG
po box 40
tuttlingen, 78501
GM  78501
Manufacturer (Section G)
AESCULAP AG
po box 40
tuttlingen, 78501
GM   78501
Manufacturer Contact
christian von der grün
po box 40
tuttlingen, 78501
GM   78501
MDR Report Key12719762
Report Number9610612-2021-00684
Device Sequence Number14416826
Product Code GDG
UDI-Device Identifier04046963909338
UDI-Public4046963909338
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/25/2024
Device Model NumberPL595SU
Device Catalogue NumberPL595SU
Device Lot Number52572608
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/21/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/05/2021
Supplement Date Received by Manufacturer05/09/2022
Initial Report FDA Received Date10/29/2021
Supplement Report FDA Received Date05/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/26/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-