Model Number OM048R
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
It was reported to aesculap (b)(4) that a roser-koenig mouth gag 160mm (part # om048r) was used during a procedure performed on an unknown date.According to the complainant, during use, the bolt holding the limiter broke.The complaint device was returned to the manufacturer for evaluation.No patient complications were reported as a result of the event.Although requested, additional information has not been made available.The malfunction is filed under aag reference (b)(4).
 
Additional Manufacturer Narrative
Manufacturing site evaluation: investigation on-going.Should relevant additional information / investigation results become available, a supplemental medwatch report will be submitted.
 
Additional Manufacturer Narrative
Based upon investigation results, this event was re-evaluated and is considered no longer reportable due to risk file - no malfunction or serious injury.Investigation results: visual investigation: the device is in a used condition, component "u-902b" is broken off.Therefore, the lock is no longer usable.Vigilance investigator carried out the pictorial documentation visually and microscopically.Component "u-902b" is broken off, fragment not available.The analysis of the fracture pattern illustrated a forced fracture due to overload.No pores, inclusions or foreign bodies could be found on the point of rupture.Due to the constant monitoring of the compliance with our quality standards, as matters stand, a production or material defect can most likely be excluded.Batch history review: the device quality and manufacturing history records (dhr) will be checked for the leading device(s) lot number(s) from the quality coordinator of the production plant.The results of the review will be documented in pc notification.If the review shows any inconspicuous, the report will be updated and actions will be initiated.Review of the complaint history revealed that no similar complaints have been filed against products from this batch number.The review of risk assessment revealed that the overall risk level (severity 1(5) x probability of occurrence (b)(4)) according to din en iso (b)(4) is still acceptable.Conclusion and measures / preventive measures: based upon the investigation results a clear root cause conclusion cannot be drawn.There is no indication for a material-, manufacturing- or design-related failure.Based upon the investigations results a capa is not necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROSER-KOENIG MOUTH GAG 160MM
Common Device Name
GENERAL SURGICAL INSTRUMENTS
Manufacturer (Section D)
AESCULAP AG
po box 40
tuttlingen, 78501
GM  78501
Manufacturer (Section G)
AESCULAP AG
po box 40
tuttlingen, 78501
GM   78501
Manufacturer Contact
christian von der grün
po box 40
tuttlingen, 78501
GM   78501
MDR Report Key12719808
Report Number9610612-2021-00685
Device Sequence Number4949675
Product Code KBN
UDI-Device Identifier04038653178940
UDI-Public4038653178940
Combination Product (Y/N)N
Initial Reporter CountryPL
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberOM048R
Device Catalogue NumberOM048R
Device Lot Number4508625409
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/14/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/05/2021
Supplement Date Received by Manufacturer01/04/2022
Initial Report FDA Received Date10/29/2021
Supplement Report FDA Received Date01/27/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/27/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-