| Model Number |
NN173Z |
| Medical Device Problem Code |
Tear, Rip or Hole in Device Packaging (2385)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/06/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Manufacturing site evaluation: investigation on-going.Should relevant additional information / investigation results become available, a supplemental medwatch report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported to aesculap ag that a as columbus ps femoral comp.Cemented f3r (part # nn173z) was implanted during a procedure performed on (b)(6), 2021.According to the complainant, the internal packaging of the femur implant was punctured which comprised the sterility of the device.Femur was autoclaved and then implanted in patient.The nurse did not open it on the field so it was contained but the surgery was delayed because we had to sterilize the femoral component in the autoclave.This event/malfunction prolonged the surgery.The complaint device was not returned to the manufacturer for evaluation.No patient complications were reported as a result of the event.Although requested, additional information has not been made available.The malfunction is filed under aag reference (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation results: as of the date of this report the complaint product was not provided for investigation.Therefore, a thorough investigation is not possible.Batch history review: the device quality and manufacturing history records have been checked for the available lot number and were found to be according to our specifications valid at the time of production.Review of the complaint history revealed that no similar complaints have been filed against products from this batch number.Conclusion and measures / preventive measures: based upon the investigation results a clear root cause conclusion cannot be drawn.There is no indication for a material-, manufacturing- or design-related failure.In the event that the complaint product will be provided for investigation in the future, an update of this report will be provided unsolicited.Based upon the investigations results a capa is not necessary.
|
| |
|
Search Alerts/Recalls
|