• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDOS INTERNATIONAL SARL SHAVER HANDPIECE 2.0 WITH BUTTONS; LINE-POWERED ARTHROSCOPIC SHAVER SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SARL SHAVER HANDPIECE 2.0 WITH BUTTONS; LINE-POWERED ARTHROSCOPIC SHAVER SYSTEM Back to Search Results
Model Number 283812
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/19/2021
Event Type  malfunction  
Manufacturer Narrative
Udi: (b)(4).To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
 
Event Description
It was reported by the sales rep that during an unknown surgery on (b)(6) 2021, it was observed that the shaver handpiece 2.0 with buttons device was short circuiting the pump.According to the report, when the device was turned on (with the foot pedal), the pump would say the software version and go into ¿stop.¿ it was reported that the pump was restarted but would turn off every time the shaver was ran.Another like device was used to complete the procedure.There were no adverse patient consequences nor surgical delay reported.No additional information was provided.
 
Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary: the complaint device is not being returned, it was retained by the customer, therefore unavailable for a physical evaluation.With the information provided, and without the complaint device to evaluate, we cannot determine a root cause for the reported failure.A manufacturing record evaluation was performed for the finished device lot number (l50ab0995), and no non-conformances related to the reported complaint condition were identified.Should the device ever be received back in the future, this complaint file will be reopened at that time and an evaluation will be performed and documented.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHAVER HANDPIECE 2.0 WITH BUTTONS
Type of Device
LINE-POWERED ARTHROSCOPIC SHAVER SYSTEM
Manufacturer (Section D)
MEDOS INTERNATIONAL SARL
chemin blanc 38
le locle CH-24 00
SZ  CH-2400
Manufacturer (Section G)
MEDOS INTERNATIONAL SARL
chemin blanc 38
le locle CH-24 00
SZ   CH-2400
Manufacturer Contact
kara ditty-bovard
325 paramount drive
raynham, MA 02767
6013142063
MDR Report Key12720460
MDR Text Key279079675
Report Number1221934-2021-03204
Device Sequence Number1
Product Code HRX
UDI-Device Identifier10886705030743
UDI-Public10886705030743
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K041824
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 10/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/29/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number283812
Device Catalogue Number283812
Is the Reporter a Health Professional? No
Date Manufacturer Received12/13/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-