The device, used in treatment, was not returned for evaluation but the pictures were reviewed, and the failure mode was confirmed.The clinical/medical investigation concluded that, smith and nephew has not received the device/adequate clinical documentation to fully evaluate the complaint.Per subsequent e-mail, the surgeon reported a 4.7 locking osteo screw pass through the evos anti-glide plate and provided two unlabeled, undated photos of the opened operative leg.Although the photos support the complaint, they don¿t aid in determining a definitive root cause of the reported failure.Therefore, the impact to the patient beyond that which has already been reported cannot be determined.Should any additional relevant medical information be provided, this complaint would be re-assessed.A review of complaint history did not reveal additional complaints for the listed device for the same failure mode.A review of the risk management file and instructions for use documents revealed this failure mode was previously identified.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Possible causes could include but not limited to surgical technique used, size of device or user/procedural variance.Based on this investigation, the need for corrective action is not indicated.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|