MAUDE Adverse Event Report: BIOTRONIK SE & CO. KG COROX OTW-S 85-BP; LV LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOTRONIK SE & CO. KG COROX OTW-S 85-BP; LV LEAD Back to Search Results
Model Number 355149
Medical Device Problem Codes Electrical /Electronic Property Problem (1198); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Pocket Erosion (2013)
Date of Event 12/11/2015
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
An infection was observed following the implantation of this biotronik device.The sterilization process was investigated.The validated process assures that all sterilization parameters, such as gas concentration, temperature, humidity, etc., are within its specified ranges for each distributed device.Additionally an analysis of validated microbiological indicators is performed after every sterilization procedure as evidence of successful completion of the sterilization process.Review of the biotronik complaint database did not reveal any changes regarding the trend for this type of incident.In summary, the infection was not device related.
 
Event or Problem Description
Lead explanted during pocket revision due to pocket erosion.Lead also had an electrical issue during explant causing the first new icd attempting to be implanted to short circuit and need to be replaced.Should additional information become available, it will be added to this file.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COROX OTW-S 85-BP
Common Device Name
LV LEAD
Manufacturer (Section D)
BIOTRONIK SE & CO. KG
woermannkehre 1
berlin 12359
DE  12359
MDR Report Key12721049
Report Number1028232-2021-05905
Device Sequence Number3706124
Product Code OJX
UDI-Device Identifier04035479094164
UDI-Public04035479094164
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P070008
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2015
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 10/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2018
Device Model Number355149
Device Catalogue NumberSEE MODEL NO.
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/26/2021
Initial Report FDA Received Date10/29/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
Patient Age78 YR
-
-