MAUDE Adverse Event Report: BALLARD CLOSED SUCTION SYSTEM FOR ADULTS, 14 F, DSE; VAP CLOSED SUCTION CATHETERS & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BALLARD CLOSED SUCTION SYSTEM FOR ADULTS, 14 F, DSE; VAP CLOSED SUCTION CATHETERS & ACCESSORIES Back to Search Results
Model Number 22103-5
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
It was reported, "the inside piece of plastic (that attaches to the trach) came unattached from the outer plastic piece (that swivels).The inner portion was stuck on the patient's trach, and took some effort to get it off." there was not reported injury.Additional information received 13-oct-2021 states the device was in place for less than a week.The device was not altered from its original form.Respiratory therapy extracted the broken piece from the bivona tts trach tube.
 
Additional Manufacturer Narrative
The actual complaint product was not returned for evaluation.A review of the device history record is not possible as no lot number was provided.All information reasonably known as of 29-oct-2021 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical, inc.Represents all of the known information at this time.Avanos medical, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical, inc.Complaint database and identified as complaint (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BALLARD CLOSED SUCTION SYSTEM FOR ADULTS, 14 F, DSE
Common Device Name
VAP CLOSED SUCTION CATHETERS & ACCESSORIES
MDR Report Key12725035
Report Number8030647-2021-00044
Device Sequence Number14449053
Product Code BSY
UDI-Device Identifier00609038983110
UDI-Public00609038983110
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 10/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number22103-5
Device Catalogue NumberN/A
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/05/2021
Initial Report FDA Received Date10/30/2021
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
BIVONA TTS TRACH TUBE
Patient Age82 YR
-
-