• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST PAUL BCI; ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST PAUL BCI; ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE Back to Search Results
Model Number 9004051
Medical Device Problem Code Device Alarm System (1012)
Health Effect - Clinical Code Insufficient Information (4580)
Type of Reportable Event Malfunction
Event or Problem Description
While calibrating it gave an occlusion error.
 
Additional Manufacturer Narrative
One unit was returned for investigation.Upon physical inspection, it was found that the complained issue could be duplicated.The failure resulted from clogged nafion tubing.A manufacturing dhr review was not performed because the results of the complaint investigation does not indicate a problem with the manufacture of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BCI
Common Device Name
ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer Contact
david halverson
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key12725431
Report Number3012307300-2021-10483
Device Sequence Number9904889
Product Code CCK
UDI-Device Identifier10610586036873
UDI-Public10610586036873
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K970209
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date (Section B) 12/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number9004051
Device Catalogue Number9004051
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/08/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/04/2021
Supplement Date Received by Manufacturer11/19/2021
Initial Report FDA Received Date10/30/2021
Supplement Report FDA Received Date12/16/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/14/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-