Model Number CA13B0/4P5/000JP
Device Problem Material Fragmentation (1261)
Patient Problems Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/01/2021
Event Type  malfunction  
Event Description
It was reported that during the pre-use check, the cushion was found torn.No patient injury.
 
Manufacturer Narrative
This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms#(b)(4).No problem or issues were identified during the device history record review.A sample was received for evaluation in an opened package.During visual inspection, the reported problem was verified.The cushion was deflated and a scratch on the cushion part was found.A v-shaped cut was found on the mask cushion.The scratches resembled those made with sharp blades.This issue was not associated with a pattern or a trend.Since the visual inspection of this product is not required in the manufacturing procedure, it is probable that it was difficult to detect this defect.Root cause was found to be a supplier item fault.The sample was sent to supplier for further investigation.D4 (udi), g3, g5 are unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX GENERAL ANESTHESIA CIRCUITS
Type of Device
BREATHING-CIRCUIT
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK  CT21 6JL
Manufacturer (Section G)
NULL
Manufacturer Contact
jim vegel
MDR Report Key12725652
MDR Text Key279259805
Report Number3012307300-2021-10490
Device Sequence Number1
Product Code CAI
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 05/04/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/31/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCA13B0/4P5/000JP
Device Catalogue NumberCA13B0/4P5/000JP
Device Lot Number210819
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/13/2021
Was the Report Sent to FDA? No
Date Manufacturer Received11/24/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-