| Model Number |
3301A1 |
| Medical Device Problem Code |
Failure to Power Up (1476)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/30/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that a smiths medical oximeter won't turn on.No adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected data: date of event updated, not used on patient, and no adverse effects.
|
| |
|
Additional Manufacturer Narrative
|
|
Other, other text: one unit was returned for investigation.Upon physical inspection, it was found that the complained issue could be duplicated.Internal inspection of the inside of the device, found power pin j9 with no solder.The loose pin j9 would cause an intermittently power issues with the device.Solder was added to pin j9.The device was powered on and the intermittent power issue was resolved.
|
| |
|
Additional Manufacturer Narrative
|
|
This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).Please disregard the initial report submitted with the mfr number: 3012307300-2021-10493.The report was submitted in error.
|
| |
|