MAUDE Adverse Event Report: BIOMET UK LTD. OXF ANAT BRG LT MD SIZE 5 PMA; OXFORD HXLPE BEARINGS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET UK LTD. OXF ANAT BRG LT MD SIZE 5 PMA; OXFORD HXLPE BEARINGS Back to Search Results
Model Number N/A
Medical Device Problem Code Patient Device Interaction Problem (4001)
Health Effect - Clinical Code Pain (1994)
Date of Event 10/18/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Initial report.Report source, foreign - event occurred in (b)(6).Customer has indicated that the product will not be returned to zimmer biomet for investigation.Multiple mdr reports were filed for this event, please see associated reports: 3002806535-2021-00472, 3002806535-2021-00474.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.Device not returned.
 
Event or Problem Description
It was reported that a patient underwent an initial knee arthroplasty on (b)(6) 2017.Subsequently, a revision due to pain was performed on (b)(6) 2021.
 
Additional Manufacturer Narrative
(b)(4).This final report is being submitted to relay additional information.Multiple mdr reports were filed for this event, please see associated reports: 3002806535-2021-00472-1.3002806535-2021-00474-1.Product has not been returned for evaluation, and x-rays or medical notes have not been provided.Therefore, the investigation has been limited to the information provided, a review of the device history records, and a complaint history search.These devices are used for treatment.No compatibility issues were noted.Review of the device history records identified no deviations or anomalies during manufacturing that could be related to the reported event.A review of complaint history was assessed for three years prior to the notification date and identified 3 similar complaints for item 154926.There were no additional complaints against the lot 3981405.A review of complaint history was assessed for three years prior to the notification date and identified 2 similar complaints for item 159549.There were no additional complaints against the lot 3850890.A review of complaint history was assessed for three years prior to the notification date and identified no additional complaints for item 166574.There were no additional complaints against the lot 3831429.These part and lot combinations are not associated with any recalls at the time the search was conducted.No corrective actions, preventive actions, or field actions resulted after investigation of this event.The likely condition of the devices when they left zimmer biomet is conforming to specification.The root cause of the reported event cannot be determined with the information provided if any additional information becomes available, then the complaint will be reopened and further investigated.
 
Event or Problem Description
It was reported that a patient underwent an initial knee arthroplasty on (b)(6) 2017.Subsequently, a revision due to pain was performed on (b)(6) 2021.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OXF ANAT BRG LT MD SIZE 5 PMA
Common Device Name
OXFORD HXLPE BEARINGS
Manufacturer (Section D)
BIOMET UK LTD.
waterton industrial estates
bridgend CF31 3XA
UK  CF31 3XA
Manufacturer (Section G)
BIOMET UK LTD.
waterton industrial estates
bridgend CF31 3XA
UK   CF31 3XA
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key12727734
Report Number3002806535-2021-00473
Device Sequence Number14587219
Product Code NRA
UDI-Device Identifier05019279785957
UDI-Public05019279785957
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
P010014
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/01/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/03/2021
Device Model NumberN/A
Device Catalogue Number159549
Device Lot Number3850890
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/22/2021
Supplement Date Received by Manufacturer11/30/2021
Initial Report FDA Received Date11/01/2021
Supplement Report FDA Received Date12/01/2021
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/03/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexUnknown
-
-