TELEFLEX MEDICAL HOL ML 5MM ENDO APPLIER; APPLIER, SURGICAL, CLIP Back to Search Results
Model Number IPN915189
Device Problem Material Separation (1562)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/12/2021
Event Type  malfunction  
Manufacturer Narrative
Qn#: (b)(4).The device has not been returned for investigation.Teleflex will continue to monitor and trend related events.
 
Event Description
The jaws were found broken during a pretest before use.Also, the pivot pin was detached.The applier were purchased by the hospital within 1 year.
 
Event Description
The jaws were found broken during a pretest before use.Also, the pivot pin was detached.The applier were purchased by the hospital within 1 year.
 
Manufacturer Narrative
(b)(4).The dhr for the returned instrument was reviewed and found completely without any irregularities.This instrument was produced at the tecomet, inc.Kenosha wi facility as part of a (b)(4) pc.Lot in june of 2020.Evaluation of the returned instrument shows that the tips are bent to one side and loose and misaligned, (pie's attached) and the jaw pivot pin is pushed thru one side of the damaged/bent outer tube assembly.We are able to validate this complaint.After the initial evaluation this instrument was dis-assembled in order to evaluate its internal components and it was found that the inner drive rod (n00185) is bent/damaged, and its bosses are both damaged where they engage the jaws.We are unable to determine how the drive rod (n00185) became bent/damaged and for the drive rod bosses to become damaged and for the jaws to become loose and misaligned and for the jaw pivot pin to be pushed thru one side of the outer tube assembly but mishandling at the end user's facility is suspected.All (b)(4) instruments from this lot were 100% visually inspected and function tested prior to shipment to the customer as this is a standardized procedure at this facility for this product line.Due to these findings, no further actions will be taken in response to this complaint and this record will be deemed closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOL ML 5MM ENDO APPLIER
Type of Device
APPLIER, SURGICAL, CLIP
Manufacturer (Section D)
TELEFLEX MEDICAL
morrisville NC
Manufacturer (Section G)
TELEFLEX MEDICAL
3015 carrington mill blvd
morrisville NC 27560
Manufacturer Contact
jasmine brown
3015 carrington mill blvd
morrisville, NC 27560
9193614124
MDR Report Key12727910
MDR Text Key279278991
Report Number3011137372-2021-00310
Device Sequence Number1
Product Code GDO
UDI-Device Identifier24026704696813
UDI-Public24026704696813
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 10/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/01/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberIPN915189
Device Catalogue Number544965
Device Lot Number06B2098690
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/04/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A