• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION GLADIATOR ELITE; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION GLADIATOR ELITE; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL Back to Search Results
Model Number 24686
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/13/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The gladiator elite 12.0 x40, 75cm, 14 atm was retuned for analysis.A visual examination identified that the balloon was not folded which indicates that the balloon was subjected to positive pressure.Blood was present in the balloon which is indicative of a leak.The rated burst pressure for this device is 14 atmospheres as per gladiator elite specification.The returned device was attached to an encore inflation unit and positive pressure was applied and liquid was observed to be leaking from a balloon pinhole located approximately 14mm proximally of the proximal marker band.An examination of the balloon material and proximal marker band identified no issues which could potentially have contributed to this complaint.A visual and microscopic examination observed no damage to the tip of the device.A visual and tactile examination of the shaft identified multiple kinks along the length of the shaft.A visual examination found no issues or damage to the markerbands of the device.
 
Event or Problem Description
Reportable based on device analysis completed on 14oct2021.It was reported that balloon leak occurred.A 12.0 x40, 75cm gladiator elite balloon catheter was selected for use.However, during preparation the balloon leaked when testing it with water.The procedure was completed with another of the same device.There were no patient complications reported.However, returned device analysis revealed a balloon pinhole.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GLADIATOR ELITE
Common Device Name
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key12727983
Report Number2134265-2021-13288
Device Sequence Number1558671
Product Code LIT
UDI-Device Identifier08714729809555
UDI-Public08714729809555
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K132810
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 11/01/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/11/2022
Device Model Number24686
Device Catalogue Number24686
Device Lot Number0023471883
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/30/2021
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/14/2021
Initial Report FDA Received Date11/01/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age44 YR
Patient Weight52
-
-