BOSTON SCIENTIFIC CORPORATION GLADIATOR ELITE; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL Back to Search Results
Model Number 24686
Device Problem Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/13/2021
Event Type  malfunction  
Manufacturer Narrative
The gladiator elite 12.0 x40, 75cm, 14 atm was retuned for analysis.A visual examination identified that the balloon was not folded which indicates that the balloon was subjected to positive pressure.Blood was present in the balloon which is indicative of a leak.The rated burst pressure for this device is 14 atmospheres as per gladiator elite specification.The returned device was attached to an encore inflation unit and positive pressure was applied and liquid was observed to be leaking from a balloon pinhole located approximately 14mm proximally of the proximal marker band.An examination of the balloon material and proximal marker band identified no issues which could potentially have contributed to this complaint.A visual and microscopic examination observed no damage to the tip of the device.A visual and tactile examination of the shaft identified multiple kinks along the length of the shaft.A visual examination found no issues or damage to the markerbands of the device.
 
Event Description
Reportable based on device analysis completed on 14oct2021.It was reported that balloon leak occurred.A 12.0 x40, 75cm gladiator elite balloon catheter was selected for use.However, during preparation the balloon leaked when testing it with water.The procedure was completed with another of the same device.There were no patient complications reported.However, returned device analysis revealed a balloon pinhole.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GLADIATOR ELITE
Type of Device
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key12727983
MDR Text Key279287323
Report Number2134265-2021-13288
Device Sequence Number1
Product Code LIT
UDI-Device Identifier08714729809555
UDI-Public08714729809555
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K132810
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Physician
Type of Report Initial
Report Date 11/01/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/11/2022
Device Model Number24686
Device Catalogue Number24686
Device Lot Number0023471883
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/30/2021
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/14/2021
Initial Date FDA Received11/01/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/12/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age44 YR
Patient Weight52
-
-