Model Number 0998-00-3023-53
Device Problems Device Displays Incorrect Message (2591); Appropriate Term/Code Not Available (3191)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/09/2021
Event Type  malfunction  
Manufacturer Narrative
Type of investigation not yet determined: additional information is being requested in regards to whether or not the customer has requested for getinge to service the iabp unit involved.A supplemental report will be submitted when additional information is provided.The full name of the initial reporter that is abbreviated is (b)(6).
 
Event Description
It was reported that the cs300 intra-aortic balloon pump (iabp) displayed a "iab disconnected" message.It is unknown the circumstances under which the event occurred.However, there was no patient involvement, and no adverse event reported.
 
Manufacturer Narrative
Device not accessible for testing (4117/3233): the getinge field service engineer called the customer to schedule service and reported that the customer's biomed cancelled the call.The fse was informed that the iabp unit had been released for clinical use.Communication/interviews (4111/3233): the customer's biomed reported that no repair was performed on the iabp unit and that the iabp unit was operational and had been cleared for clinical use.A supplemental report will be submitted upon completion of our investigation.H3 other text : not returned to manufacturer.
 
Event Description
N/a.
 
Manufacturer Narrative
Updated fields: b4, g4, g7, g8, h2, h4, h6(investigation type, investigation findings & investigation conclusions), h10, h11.Corrected fields: d5, g1(contact person).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key12728179
MDR Text Key279302130
Report Number2249723-2021-02516
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567107882
UDI-Public10607567107882
Combination Product (y/n)N
PMA/PMN Number
K063525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup,Followup
Report Date 09/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/01/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0998-00-3023-53
Device Catalogue Number0998-00-3023-53
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/01/2022
Was Device Evaluated by Manufacturer? No
Date Device Manufactured05/15/2007
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-