Model Number DR8084
Medical Device Problem Code Packaging Problem (3007)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/09/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Manufacturing review: a manufacturing review was not requested as the lot number reported as unknown.Investigation summary: the physical device was not returned for evaluation.Two electronic photos were reviewed.The first photo shows the shipping carton which appears to be bent and damaged at its midsection.The second photo shows the outer packaging box of the dorado pta balloon dilatation catheter.The midsection of the packaging box appears to be crushed and bent.No evidence of sterile breach noted on the device.Therefore, based on the photo review, the reported packaging problem can be confirmed as the boxes appear to be damaged and bent.A definitive root cause for the alleged packaging problem could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.
 
Event or Problem Description
It was reported that prior to an angioplasty procedure, the package of the device was allegedly damaged.There was no patient contact.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DORADO
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key12728550
Report Number2020394-2021-01928
Device Sequence Number1467404
Product Code LIT
UDI-Device Identifier00801741059254
UDI-Public(01)00801741059254
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K072283
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/06/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberDR8084
Device Catalogue NumberDR8084
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/04/2021
Initial Report FDA Received Date11/01/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-