U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER ORTHOPAEDIC MFG. LTD. FEMUR CEMENTED (PS) STANDARD RIGHT SIZE 7; PROSTHESIS, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER ORTHOPAEDIC MFG. LTD. FEMUR CEMENTED (PS) STANDARD RIGHT SIZE 7; PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Pain (1994); Loss of Range of Motion (2032); Ambulation or Postural Difficulties (2544); Swelling/ Edema (4577)
Date of Event 07/16/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Multiple mdr reports were filed for this event, please see associated report: 0002648920-2021-00383, 0002648920-2021-00384.Patient¿s dob: (b)(6) 1947.Medical product tibia cemented 5 degree stemmed right size e item# 42532007102 lot# 63180918.Articular surface fixed bearing (ps) right 11 mm height item# 42521400711 lot# 62816413.Report source- (b)(6).Customer has indicated that the product will not be returned to zimmer biomet for investigation as it remains implanted.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that a patient enrolled in a clinical study experienced severe stiffness and difficulty with activities of daily living four years post implantation.At the five year follow up patient reported severe pain.Patient is confined to bed due to difficulty with mobility and swelling.No interventions have been reported to date.Attempts to obtain additional information have been made; however, no more information is available at this time.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up is being submitted to relay additional information.No product was returned, or pictures provided; therefore, visual and dimensional evaluations could not be performed.Medical records were provided and reviewed by a health care professional.Review of the available records identified the following: four year follow up: pain 3/10, moderate-severe difficulty with dls, no swelling, grinding/clicking often, severe stiffness, no complications on x-ray; five year follow up: no instability, rom to 90 degree flexion, pain 10/10, confined to bed, swelling always present, grinding/clicking rare, moderate stiffness, extreme difficulties, no complications on x-ray.Device history record (dhr) was reviewed for deviations and/ or anomalies with no deviations / anomalies identified.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FEMUR CEMENTED (PS) STANDARD RIGHT SIZE 7
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
EI 
Manufacturer (Section G)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
EI  
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key12729563
Report Number3007963827-2021-00266
Device Sequence Number4537880
Product Code JWH
UDI-Device Identifier00889024229228
UDI-Public(01)00889024229228(17)250731(10)63131009
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K113369
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/01/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/31/2025
Device Model NumberN/A
Device Catalogue Number42500606202
Device Lot Number63131009
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/24/2021
Supplement Date Received by Manufacturer11/09/2021
Initial Report FDA Received Date11/01/2021
Supplement Report FDA Received Date12/01/2021
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10
Outcome Attributed to Adverse Event Other;
Patient Age74 YR
Patient SexFemale
Patient Weight95 KG
-
-