• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EDWARDS LIFESCIENCES FEMEROL VENOUS CANNULA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES FEMEROL VENOUS CANNULA Back to Search Results
Model Number VFEM022
Medical Device Problem Code Packaging Problem (3007)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/07/2021
Type of Reportable Event Malfunction
Event or Problem Description
Edwards was notified that three cannulas were received in wet cartons with signs of moisture inside the sterile package.
 
Additional Manufacturer Narrative
Any packaging non-conformance that results in a breach of the sterile package (i.E.Pinhole, tear or seal issue), has the potential to result in a serious infection.The device is in the process of being returned.The investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.
 
Additional Manufacturer Narrative
H3: product evaluation: customer report of moisture inside the sterile package was not confirmed in any of the three samples.Sample 1 as received no moisture or signs of moisture were noticed, however the left area of outer carton box was noticed curvy and inner layer of the carton box lid was found detached.Inner tab was also found curvy and folded inward.Inner packing was free of any visible damage or moisture.Sample 2 as received no moisture or signs of moisture were noticed, however the left area of outer carton box was noticed curvy and inner layer of the carton box lid was found detached.Inner packing was free of any visible damage or moisture.Sample 3 as received no moisture or signs of moisture were noticed, however from middle to the right area of outer carton box the box lid was noticed curvy.Inner packing was free of any visible damage or moisture.All three returned units appeared consistent with customer photos, however water drops were not visible.
 
Additional Manufacturer Narrative
H10: additional narratives.Updated d4, h4, and h6 per new information received.The root cause of this event cannot be conclusively determined.However, it is likely that the units were exposed to moisture during storage/transportation in route to the customer.The device history record (dhr) was reviewed and showed that this device met all manufacturing and sterilization specifications for product release prior to distribution.No issues were identified that would have impacted this event.Edwards will continue to review and monitor all events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FEMEROL VENOUS CANNULA
Common Device Name
FEMEROL VENOUS CANNULA
Manufacturer (Section D)
EDWARDS LIFESCIENCES
one edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES
one edwards way
irvine CA 92614
Manufacturer Contact
gail warner
one edwards way
mailstop 6.1 anton
irvine, CA 92614
9492504096
MDR Report Key12729723
Report Number2015691-2021-06103
Device Sequence Number1479875
Product Code DWF
UDI-Device Identifier00690103146660
UDI-Public(01)00690103146660(17)230811(11)2008112163331136
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K891576
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Expiration Date08/11/2023
Device Model NumberVFEM022
Device Catalogue NumberVFEM022
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/09/2021
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/07/2021
Supplement Date Received by Manufacturer11/18/2021
12/23/2021
Initial Report FDA Received Date11/01/2021
Supplement Report FDA Received Date11/30/2021
01/03/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/11/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-