MAUDE Adverse Event Report: PHILIPS IMAGE GUIDED THERAPY CORPORATION ANGIOSCULPT EVO RX PTCA; CATHETER, PTCA, CUTTING/SCORING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS IMAGE GUIDED THERAPY CORPORATION ANGIOSCULPT EVO RX PTCA; CATHETER, PTCA, CUTTING/SCORING Back to Search Results
Model Number 2200-3015-B
Medical Device Problem Codes Entrapment of Device (1212); Improper or Incorrect Procedure or Method (2017)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/06/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The patient's ethnicity and race are unknown.This information was not available from the facility.Patient information regarding relevant tests/laboratory data or medical history are unknown.This information was not available from the facility.Attempts to obtain the information has not been successful.The angiosculpt device was returned for evaluation.Visual inspection found a distal bond peel with one leaflet lifted, but remained intact to the device.During functional testing, a laboratory asahi sion blue 0.014" coated guide wire was activated with saline and inserted through the distal tip and through the device, with slight resistance.After balloon inflation/deflation, with slight resistance the guide wire was removed from the device.The angiosculpt device was used off-label.The ifu warns to not exceed the rated burst pressure as indicated on the product label.
 
Event or Problem Description
The angiosculpt device was used to treat the proximal rca and upon removal, the balloon got pinned to the wire and had to be removed as a unit.The balloon was inflated 3 times to 18 atm (rbp is 16 atm).No issues observed or reported during delivery.The lesion was rewired without complications to complete the procedure.No patient injury reported.During the returned device analysis, a distal bond peel was observed.This adverse event is being submitted because the angiosculpt and guidewire were removed as a unit.Placement of a new guide wire was necessary to complete the procedure.In addition, this product problem is being submitted because the distal bond peeled, potential for harm if it were to recur.
 
Additional Manufacturer Narrative
Blocks d1/ g1: manufacturer name and site name were updated from spectranetics to philips image guided therapy corporation to match the information on the product label.Submission of this report does not, in itself, represent a conclusion by the manufacturer and/or authorized representative or the national competent authority that the content of this report is complete or accurate, that the medical device(s) listed failed in any manner and/or that the medical device(s) caused or contributed to an alleged death or deterioration in the state of the health of any person.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANGIOSCULPT EVO RX PTCA
Common Device Name
CATHETER, PTCA, CUTTING/SCORING
Manufacturer (Section D)
PHILIPS IMAGE GUIDED THERAPY CORPORATION
5055 brandin court
fremont CA 94538
Manufacturer (Section G)
PHILIPS IMAGE GUIDED THERAPY CORPORATION
5055 brandin court
fremont CA 94538
Manufacturer Contact
ana tan
5055 brandin court
fremont, CA 94538
MDR Report Key12733990
Report Number3005462046-2021-00061
Device Sequence Number14587220
Product Code NWX
UDI-Device Identifier00813132026899
UDI-Public00813132026899
Combination Product (Y/N)Y
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
P050018
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/02/2022
Device Model Number2200-3015-B
Device Catalogue Number2200-3015-B
Device Lot NumberG21030006
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/19/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/06/2021
Supplement Date Received by Manufacturer10/06/2021
Initial Report FDA Received Date11/01/2021
Supplement Report FDA Received Date08/08/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/02/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ASAHI: 0.014" SION BLUE GUIDEWIRE; BOSTON SCIENTIFIC: AL GUIDE CATHETER SIZE UNK; TERUMO: 6F INTRODUCER SHEATH
Patient Age54 YR
Patient SexFemale
Patient Weight74 KG
-
-