• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. RETROGRADE DRL 9.5MM; BIT, SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. RETROGRADE DRL 9.5MM; BIT, SURGICAL Back to Search Results
Model Number 72204045
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/30/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event or Problem Description
It was reported that during use in an anterior cruciate ligament surgery, the "retrograde drl" broke in during the drilling.The procedure was completed with a delay greater than 30 minutes and was finished with a smith and nephew back up device.No patient complications were reported.
 
Additional Manufacturer Narrative
The reported device was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.There was no way to determine if the device contributed to the reported event.A complaint history review concluded this was a repeat issue.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The complaint was not confirmed.Factors that could have contributed to the reported event include an application of unintended inappropriate or excessive force to the device, attempted correction of a damaged device, or an impact event inconsistent with normal use.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RETROGRADE DRL 9.5MM
Common Device Name
BIT, SURGICAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key12734202
Report Number1219602-2021-02409
Device Sequence Number16504413
Product Code GFG
UDI-Device Identifier00885554037746
UDI-Public00885554037746
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Consumer,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/18/2024
Device Model Number72204045
Device Catalogue Number72204045
Device Lot Number4937193
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/11/2021
Supplement Date Received by Manufacturer02/24/2022
Initial Report FDA Received Date11/01/2021
Supplement Report FDA Received Date03/03/2022
Date Device Manufactured10/18/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-