BOSTON SCIENTIFIC CORPORATION INTUA; PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P) Back to Search Results
Model Number V273
Device Problem Premature Discharge of Battery (1057)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/23/2021
Event Type  Injury  
Manufacturer Narrative
Upon receipt at our post market quality assurance laboratory, a thorough evaluation of the device was performed.The allegation premature battery depletion (pbd) was not confirmed.The baseline testing completed on the device found no failing conditions.Further, all testing completed on the device passed or was not applicable, therefore no problem was detected.
 
Event Description
It was reported that this cardiac resynchronization therapy pacemaker (crt-p) was explanted due to premature battery depletion (pbd).A new device was implanted.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTUA
Type of Device
PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key12735828
MDR Text Key279527511
Report Number2124215-2021-33293
Device Sequence Number1
Product Code NKE
UDI-Device Identifier00802526536670
UDI-Public00802526536670
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030005/S092
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 11/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/02/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date02/26/2017
Device Model NumberV273
Device Catalogue NumberV273
Device Lot Number118675
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/28/2021
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/23/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/26/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age74 YR
-
-